For more quantitative results than phase detection or Amplitude Detection, use frequency modulation (FM). This technique provides a direct measure of the change in resonant frequency sensed by the cantilever.
Prior to initiating this procedure you must have completed the steps outlined in Basic EFM Operation.
) = 40 and Input Pgain (proportional gain
) = 60.